A Driven Right Leg circuit or DRL circuit, also known as Right Leg Driving technique, is an electric circuit that is often added to biological signal amplifiers to reduce common-mode interference. Biological signal amplifiers such as ECG (electrocardiogram) EEG (electroencephalogram) or EMG circuits measure very small electrical signals emitted by the body, often as small as several micro-volts. However, the patient's body can also act as an antenna which picks up electromagnetic interference, e